Journal of Scientific and Technical Research

Abstract

This article presents an improved configuration for fitting the radial dose functions of sealed 192Ir #brachytherapy sources. The proposed functions allow for precise fitting of #radial dose function in the range of 0.5 20 cm. The maximum deviations between fitting data and Monte Carlo modeling data do not exceed 0.6%, while average deviations are in the order of 0.1%. The fitting coefficients for the proposed functions are specified for 11 sources from different manufacturers. The proposed functions could be used in #treatment planning systems for more accurate calculation of the dose distributions around the sources using a wide range of data according to TG-43 #dosimetry protocol.

For more articles on BJSTR Journals please click here: https://biomedres.us/