Wednesday, October 31, 2018

Wear in Two Cases of Ceramic on #Ceramic Total #HipArthroplasty After 30 Years Follow Up by Torres A in BJSTR

Abstract

Background: #Ceramic-on-ceramic (CoC) bearings have been in use in total #hipreplacement (THR) for more than 40 years, with excellent long-term survivorship but concerns over the ceramic-on-ceramic bearings wear persist.There are few studies where they analysed the ceramic-on-ceramic #implants retrieved. We wanted to investigate the wear patterns and the surface properties of the CoC first generation THR implants removed.
Case description: We have analysed two cases of ceramic head on ceramic liner of Autophor primary THA implanted in active patients and retrieved for aseptic loosening after 34 and 31 years of well function.The sphericity, surface roughness and #microtopography was assess by a profilometer and a white light interferometer. Scanning #electronmicroscopy of the implant-retrieved surfaces was also perform.




No comments:

Post a Comment

Note: Only a member of this blog may post a comment.

Centrifugal Pneumatic-Sieve Fractionation of Barley and Wheat Flour for Maximum Recovery of Protein and Carbohydrate Fractions

  Centrifugal Pneumatic-Sieve Fractionation of Barley and Wheat Flour for Maximum Recovery of Protein and Carbohydrate Fractions Introductio...